Klausen Trading Company Pty Ltd

Erweka - RoboDis Technical Specifications

 

HARDWARE

Filling of Vessels 500ml – 1000ml, ±1% (gravimetric filling)
Stirring Speed 25 rpm – 200 rpm, ± 1 rpm
Precision of
Temperature Control
± 0.1°C
Temperature in Vessels 37°C, ± 0.5°C
Precision of
Sampling Time
± 4%
Cross-contamination ± 1% (with cleaning of optrode)
Reproducibility of Results VC ± 1%
Sampling Times
with Optrode
± 5 min. for 6 test stations (with cleaning of optrode)
± 3 min. for 6 test stations (without cleaning of optrode)
Evaporation from Vessel ± 1% in 24h

 

SOFTWARE
 

Operation System MS Windows XP / Vista
Audit Trail According to 21 CFR Part 11
User Management Detailed allocation of user rights possible
External Interfaces

Export of analytical results and measurements into different formats (e.g. CSV)

  • Communication with external analytical instruments (HPLC, UV-VIS)
  • LIMS interface
Software Validation validation documents included in delivery, PQ documents available