Erweka - RoboDis Technical Specifications
HARDWARE
| Filling of Vessels | 500ml – 1000ml, ±1% (gravimetric filling) |
| Stirring Speed | 25 rpm – 200 rpm, ± 1 rpm |
| Precision of Temperature Control |
± 0.1°C |
| Temperature in Vessels | 37°C, ± 0.5°C |
| Precision of Sampling Time |
± 4% |
| Cross-contamination | ± 1% (with cleaning of optrode) |
| Reproducibility of Results | VC ± 1% |
| Sampling Times with Optrode |
± 5 min. for 6 test stations (with cleaning of optrode) ± 3 min. for 6 test stations (without cleaning of optrode) |
| Evaporation from Vessel | ± 1% in 24h |
SOFTWARE
| Operation System | MS Windows XP / Vista |
| Audit Trail | According to 21 CFR Part 11 |
| User Management | Detailed allocation of user rights possible |
| External Interfaces |
Export of analytical results and measurements into different formats (e.g. CSV)
|
| Software Validation | validation documents included in delivery, PQ documents available |